Siemens Healthcare Diagnostics, Inc. — FDA Class II Enforcement Record
- Report Date
- 2026-11-18
- Initiation Date
- 2026-09-22
- Status
- Ongoing
- Firm Location
- Tarrytown, NY, United States
Reason for FDA Recall
Reagent probe carryover on the Atellica CH and Atellica CI Analyzers may cause falsely elevated Hemoglobin A1c (A1c_E/A1c_H) results due to HDLC carryover (biases of 9 29% in NGSP units), and falsely depressed Direct Bilirubin 2 (DBil_2) results due to TP_2 carryover (biases of 0% to -25%), which may affect patient samples, quality control, and/or calibrator results.
Affected Product Description
Atellica CH Total Protein_2 (TP_2); Siemens Material Number (SMN): 1. 11645755 (3700 tests), 2. 11537228 (7400 tests);
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Burkina Faso, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cura¿ao, Cura¿ao, St Eus, Czech Republic, Democratic…