Class II Medical & Beauty Devices Recall #Z-0379-2027

Vesco Medical Llc — FDA Class II Enforcement Record

Report Date
2026-11-18
Initiation Date
2026-10-08
Status
Ongoing
Firm Location
Westerville, OH, United States

Reason for FDA Recall

The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences.

Affected Product Description

Vesco Q 100mL Small Bore Bag Feed Set, Model Number VED-059

Distribution Pattern

US Nationwide distribution.

Risk Classification Tags

Misbranding / Label Defect

Activate Automated FDA Monitoring or Bulk Pass

Watchlist Radar Pro ($29/mo)

Format: you@company.com (Used strictly for License Key delivery & FDA alerts)