TransMedics, Inc. — FDA Class II Enforcement Record
- Report Date
- 2026-11-18
- Initiation Date
- 2025-11-20
- Status
- Ongoing
- Firm Location
- Andover, MA, United States
Reason for FDA Recall
Faulty Solution Delivery System (SDS) Cassettes within the device set were assembled with syringes that were manufactured with an atypical lubricant, which results in sub-par performance and has the potential to cause leaks.
Affected Product Description
OCS Liver System for organ transport OCS Solution Delivery Cassettes Model/Catalog Number: 1467 Product Description: OCS Solution Delivery Cassettes The recalled product is a standard off the shelf syringe. It is used as part of solution delivery cassette (SDS Cassette) to be used with the perfusion module of OCS Liver System. Each OCS Solution Delivery Cassettes contains three SDS cassettes. The BD syringe is a component of the SDS Cassette PN 100004459
Distribution Pattern
Worldwide - US Nationwide distribution in the state of NE and the country of Italy.