Class I Medical & Beauty Devices Recall #Z-0409-2027
3B Medical, Inc. — FDA Class I Enforcement Record
- Report Date
- 2026-11-18
- Initiation Date
- 2026-10-08
- Status
- Ongoing
- Firm Location
- Dublin, OH, United States
Reason for FDA Recall
The configured IPAP value is recorded and transmitted instead of the actual delivered inspiratory pressure when Targeted VT is enabled.
Affected Product Description
Luna G3 BPAP System Model/Catalog Number: LG3800-30VT Product Description: The Luna G3 BPAP System is a Bi-level PAP (Bi-level Positive Airway Pressure) device designed for the treatment of adult Obstructive Sleep Apnea (OSA).
Distribution Pattern
US nationwide distribution including Puerto Rico. No OUS distribution.
Risk Classification Tags
Regulatory Enforcement