Class I Medical & Beauty Devices Recall #Z-0409-2027

3B Medical, Inc. — FDA Class I Enforcement Record

Report Date
2026-11-18
Initiation Date
2026-10-08
Status
Ongoing
Firm Location
Dublin, OH, United States

Reason for FDA Recall

The configured IPAP value is recorded and transmitted instead of the actual delivered inspiratory pressure when Targeted VT is enabled.

Affected Product Description

Luna G3 BPAP System Model/Catalog Number: LG3800-30VT Product Description: The Luna G3 BPAP System is a Bi-level PAP (Bi-level Positive Airway Pressure) device designed for the treatment of adult Obstructive Sleep Apnea (OSA).

Distribution Pattern

US nationwide distribution including Puerto Rico. No OUS distribution.